Important Info
Overview
The industry's first water level data logger designed for convenient wireless setup and download via Bluetooth technology. Dramatically simplifies and lowers the cost of field data collection by providing wireless access to high-accuracy water level and temperature data right from a mobile phone, tablet, or Windows computer running Onset's free HOBOconnect app. Consists of a top-end unit and a water level sensor which are sold as a set. A direct-read cable is required to connect them. Cables can be ordered in lengths from 0.2 to 500m for deployment in a wide range of wells. Add the cable from Compatible Items below or visit the Direct Read Cable product page.
Highlighted Features
- Wireless data offload to mobile devices using Bluetooth Low Energy
- Integrated barometric pressure sensor enables direct water level readout
- Direct read cable connects sensor to top-end logger/transmitter
- The cable includes Kevlar strength member
- Cables are interchangeable so loggers are easy to redeploy in future applications
- The logger and sensor add 0.39 meters to the length of the cable
- Cable length can vary up to 3% from the length ordered
- The 5, 10, 15, 30 and 60 meter cable lengths are in stock; custom cable lengths have a 2 to 4 week lead time
- Reference water level can be entered at the start of the deployment
- Use HOBOmobile‚ for setup, data viewing and data sharing‚
- Powered by two user-replaceable AA batteries in the top-end‚ unit
- Several logging modes: normal, multi-rate logging and event-triggered‚ burst-logging
- Durable ceramic sensor
- Available with stainless steel or titanium sensor ends
- 3-point NIST-traceable calibration certificate included for the water pressure sensor
Specifications
Pressure (Absolute) and Water Level Measurements MX2001-01-SS-S and MX2001-01-Ti-S - 9 Meter (30') range |
Operation Range |
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F) |
Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth |
Water Level Accuracy* |
Typical error: ±0.05% FS, 0.5 cm (0.015 ft) water Maximum error: ±0.1% FS, 1.0 cm (0.03 ft) water |
Raw Pressure Accuracy** |
±0.3% FS, 0.62 kPa (0.09 psi) maximum error |
Resolution |
<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water |
Pressure Response Time (90%)*** |
1 second at a stable temperature |
Pressure (Absolute) and Water Level Measurements MX2001-02-SS-S - 30 Meter (100') range |
Operation Range |
0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F) |
Burst Pressure |
500 kPa (72.5 psia) or 40.8 m (134 ft) depth |
Water Level Accuracy* |
Typical error: ±0.05% FS, 1.5 cm (0.05 ft) water Maximum error: ±0.1% FS, 3.0 cm (0.1 ft) water |
Raw Pressure Accuracy** |
±0.3% FS, 1.20 kPa (0.17 psi) maximum error |
Resolution |
0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water |
Pressure Response Time (90%)*** |
1 second at a stable temperature |
Pressure (Absolute) and Water Level Measurements MX2001-03-SS-S - 76 Meter (250') range |
Operation Range |
0 to 850 kPa (0 to 123.3 psia); approximately 0 to 76.5 m (0 to 251 ft) of water depth at sea level, or 0 to 79.5 m (0 to 262 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 850 kPa (10 to 123.3 psia), 0° to 40°C (32° to 104°F) |
Burst Pressure |
1,200 kPa (174 psia) or 112 m (368 ft) depth |
Water Level Accuracy* |
Typical error: ±0.05% FS, 3.8 cm (0.125 ft) water Maximum error: ±0.1% FS, 7.6 cm (0.25 ft) water |
Raw Pressure Accuracy** |
±0.3% FS, 2.55 kPa (0.37 psi) maximum error |
Resolution |
<0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) water |
Pressure Response Time (90%)*** |
1 second at a stable temperature |
Pressure (Absolute) and Water Level Measurements MX2001-04-SS-S and MX2001-04-Ti-S - 4 Meter (13') range |
Operation Range |
0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F) |
Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth |
Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water |
Raw Pressure Accuracy** |
±0.3% FS, 0.43 kPa (0.063 psi) maximum error |
Resolution |
<0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water |
Pressure Response Time (90%)*** |
<1 second at a stable temperature |
Barometric Pressure (MX2001-TOP) |
Operation and Calibrated Range |
66 to 107 kPa (9.57 to 15.52 psia); -20° to 50°C (-4° to 122°F) |
Accuracy |
±0.2 kPa (±0.029 psi) over full temperature range at fixed pressure; maximum error ±0.5% FS |
Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm (0.01 ft) water Maximum error: ±0.15% FS, 0.6 cm (0.02 ft) water |
Resolution |
0.01 kPa (0.0015 psi) |
Response Time |
1 second at stable temperature |
Stability (Drift) |
0.01 kPa (0.0015 psi) per year |
Temperature Measurements (All Sensor End Models MX2001-0x-SS-S and MX2001-0x-Ti-S) |
Operation Range |
-20° to 50°C (-4° to 122°F) |
Accuracy |
±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A |
Resolution |
0.1°C at 25°C (0.18°F at 77°F), see Plot A in manual |
Response Time (90%) |
5 minutes in water (typical) |
Stability (Drift) |
0.1°C (0.18°F) per year |
Logger |
Operation Range |
-20° to 50°C (-4° to 122°F) |
Radio Power |
1 mW (0 dBm) |
Transmission Range |
Approximately 30.5 m (100 ft) line-of-sight |
Wireless Data Standard |
Bluetooth Low Energy (Bluetooth Smart) |
Logging Rate |
1 second to 18 hours |
Logging Modes |
Fixed interval, multiple intervals with up to 8 user-defined logging intervals and durations, or event-triggered burst |
Memory Modes |
Wrap when full or stop when full |
Start Modes |
Immediate, date & time, or next interval |
Stop Modes |
When memory full, stop with HOBOmobile, date & time, or after a set logging period |
Time Accuracy |
±1 minute per month 0° to 50°C (32° to 122°F) |
Battery |
Two AA, 1.5 V alkaline batteries, user-replaceable |
Battery Life |
1 year, typical with logging interval of 1 minute. Faster logging and/or statistics sampling intervals, entering burst logging mode, excessive readouts, checking of Full Status Details, and remaining connected with HOBOmobile will impact battery life. |
Memory |
256 KB memory (30,000 sets of measurements) |
Full Memory Download Time |
Approximately 2 minutes; may take longer the further the device is from the top end of the logger |
Dimensions |
Top end (MX2001-TOP): 2.54 cm (1.0 inches) diameter, 28.9 cm (11.4 inches) length; mounting hole 7.6 mm (0.3 inches) diameter Sensor end (MX2001-0x-SS-S and MX2001-0x-Ti-S): 2.54 cm (1.0 inches) diameter, 9.91 cm (3.9 inches) length Note: The length of the water level logger cable (CABLE-DR-xxx) can vary -0% to +3% +10 cm (3.9 inches) from the length ordered. The logger adds 38.8 cm (15.3 inches) to the length of the cable ordered. |
Weight |
Top end (MX2001-TOP): Approximately 136 g (4.78 oz) in air Stainless sensor end (MX2001-0x-SS-S): Approximately 141.4 g (4.99 oz) in air; approximately 53.9 g (1.9 oz) in fresh water Titanium sensor end (MX2001-0x-Ti-S): Approximately 80 g (2.83 oz) in air; approximately 37 g (1.3 oz) in fresh water |
Wetted Materials |
Top end (MX2001-TOP): PVC housing, Polycarbonate end cap; Viton O-ring Stainless sensor end (MX2001-0x-SS-S): Stainless steel housing, Viton and Buna-N O-rings; ceramic sensor in stainless steel end cap Titanium sensor end (MX2001-0x-Ti-S): Acetal housing, Viton and Buna-N O-rings; ceramic sensor in Titanium end cap Cable (CABLE-DR-XXX): Polycarbonate strain relief (top end connection), PVC end cap (sensor connection), nylon collar nuts, Viton O-rings, polyurethane jacket |
Environmental Rating |
Top end: Weatherproof housing - NEMA 6, IP67 Sensor end: IP68 |
|
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). |
|
|
*Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected sensor end Water Level Accuracy. **Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors. ***Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes. Measurement accuracy also depends on temperature response time.
Compatible Items
MAINTENANCE Replacement Parts
- DESICCANT2 Large Desiccant Pack (25-count) Replacement
- DESICCANT1 Small Desiccant Pack (25-count) Replacement
- The MX2001 uses one small and one large desiccant pack. Because the logger works with a separate barometric sensor, it doesn't have a vent tube and frequent desiccant replacements are not required. When installing new batteries, check the desiccant and replace if necessary.
ACCESSORIES‚ Communication Cables
Well Caps